XZR500 User’s Manual
34
97137 Issue 8, April 2017
MAINTENANCE
4 MAINTENANCE
4.1
Preventative Maintenance - Cleaning
The XZR500 should be periodically cleaned. The frequency of cleaning will depend
upon your application.
•
Identify the 3 orientation marks on the inner tube and remove it (
Figure
4 (23)).
•
Clean the inside of the outer tube (
Figure 4 (2)) by moving the inner tube
backwards and forwards - the tip disk on the inner tube will sweep the
inside of the outer tube.
•
Clean the inner tube with a long metal brush or abrasive cloth. It should
then be cleaned with hot water and afterwards have pressurized air blown
through the tube. Fit the inner tube of the probe back into the correct
position.
•
Change the gaskets and tighten-up all the nuts frequently to prevent
contamination from outside air entering the stack.
•
The probe should be removed at least once a year, e.g. during a factory
shutdown, to inspect the probe tubes. Clean them as necessary.