XZR500 User’s Manual
26
97137 Issue 8, April 2017
OPERATION
Output Signal
0 or 4-20 mA
4.00
This page shows the available current ranges
for the signal output. Enter 0.00 for 0-20 mA
and 4.00 for 4-20 mA. Press the [+] and [–]
keys to adjust the value. Press the [*]
Enter
key to con
fi
rm the selection and move to the
next parameter.
Measurement range – lower limit
Rec. range low
0.01
This page shows the selected lower limit of the
measurement range. Press the [+] and [–] keys
to adjust the value. Press the [*]
Enter
key
to con
fi
rm the selection and move to the next
parameter.
NOTE: This value is in %. For example,
0.01 equals 0.01% O
2
.
Measurement range – upper limit
Rec. range up
10.00
This page shows the selected upper limit of the
measurement range. Press the [+] and [–] keys
to adjust the value. Press the [*]
Enter
key
to con
fi
rm the selection and move to the next
parameter.
NOTE: This value is in %. For example,
10.00 equals 10% O
2
.
CO
2
factor
CO
2
factor
0.00
This page is used to enter the CO
2
maximum
theoretical value of the fuel gas for CO
2
calculation. Press the [+] and [–] keys to adjust
the value. Press the [*]
Enter
key to con
fi
rm
the selection and move to the next parameter.
(See Appendix C)
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...