XZR500 User’s Manual
10
97137 Issue 8, April 2017
INSTALLATION
2.2
Probe Mechanical Installation
f. Inner probe tube
3 marks to indicate
the bevelled edge
Horizontal line
Bevelled edge
e. Probe head
d. Gasket for flange
c. Outer tube
b. Gasket for flange
a. Tubular
counter flange
Gas Inlet hole
}
}}
5°
}
}
}
}
}
}
}
h. Locking ring
i. Rear nut
g. Gasket for locking ring
}
}
}
}
}
Gasket
Nut
}
}
j. Gasket
Figure 8
Probe Installation
The XZR500 is simple to set-up. Follow the instructions below:
1.
Weld the tubular counter-
fl
ange (a) onto the stack. Follow the orientation
shown below to ensure that the probe head is set in a vertical position.
Slope the tubular counter-
fl
ange (a) slightly (maximum 5 degrees) so that
condensed water can go back to the process. The probe is
fi
tted with a PN6
DN15 type
fl
ange (4 x 11mm diameter holes, placed on a 55mm diameter
circle). Pay special attention to the orientation and slope especially if the
counter
fl
ange/nozzle with
fl
ange is provided by the customer.
055mm
055mm
055mm
9
8
8
5°
Figure 9
Probe Head Orientation
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...