XZR500 User’s Manual
14
97137 Issue 8, April 2017
INSTALLATION
2.4.2.1 Connection to the Mains
Use shielded cable (2 x 1.5mm² max.) that is terminated appropriately. Follow the
Table 4 wiring diagram (pins 48, 49 and 50).
Connect both the screen and the earth wire to the copper bar near the cable gland.
2.4.2.2 Connection of the 0/4-20 Output
We suggest that shielded cable (2 x 0.75mm²) is used and terminated appropriately.
Follow the Table 4 wiring diagram (pins 34 and 35).
Connect the braid to the copper bar near the cable gland.
2.4.2.3 Connection of the Alarms
We suggest that shielded cable is used and terminated appropriately. Follow the Table
4 wiring diagram (pins 41 to 46).
Connect the screen and the earth wire to the copper bar near the cable gland.
2.4.3 Connection of the Cable to the Sensor Head
Unscrew the three hex head screws at locations shown below to open the case of the
XZR500 probe head and access the terminal block.
Figure 13
Hex Screw Locations