XZR500 User’s Manual
22
97137 Issue 8, April 2017
OPERATION
3.3.2.1 Access code 0.12 - quick settings
Your selection?
Visu * Cal + Set -
From the main menu page, press and hold the
[*]
Enter
key until the main selection menu
appears.
Access code?
0.12
Press the [–] (Set-up menu) from the main
selection menu to show the Access code page.
Using the [–] and [+] keys enter the 0.12 code
to access the quick settings and press the [*]
Enter
key to con
fi
rm the selection.
Standard Con
fi
guration
Std con
fi
g. 0.30
0.00
The
fi
rst page under access code 0.12 is
for restoring the factory default standard
con
fi
gurations.
A standard con
fi
guration should be restored after replacing the microcontroller
card.
Consequences: All the parameters change to the default value. The analyzer
is no longer calibrated.
DO NOT change the value from 0.00 without consulting the
factory. Restoring to standard con
fi
guration will change all the
parameters of the last set-up. Make sure you note down all
parameter values before executing the operation. After re-set
the analyzer is no longer calibrated.
If a standard con
fi
guration is not necessary leave the value as 0.00 and press the [*]
Enter
key to go to the next page.
If a standard con
fi
guration is necessary, enter 0.30 and press the [*]
Enter
key to
activate the restoring process.
Calibration gas value
Test gas value
8.00
This page shows the current calibration gas
value. Use the [+] and [–] keys to modify the
value and press the [*]
Enter
key to con
fi
rm the
selection and move to the next parameter.
NOTE: This value is in %. For example,
8.00 equals 8%.
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...