OPERATION
3 OPERATION
The MSRS is a fragile element - keep the sensor free from any
shock. Any measurements that need to be done at the MSRS
terminals must be done very carefully in order to prevent
irreversible damage to the sensor.
NEVER attempt to measure the resistance between the MSRS
reference terminal and another terminal
Use a voltmeter with impedance >1000 M
Ω
for measuring
the voltage between the MSRS measurement and reference
terminals
3.1 Outputs
3.1.1 Analog
Output
The output signal is selectable to be either 0-20 mA or 4-20 mA.
The range is selectable between 0 and 25%.
3.1.2 Alarms
The system has 3 alarm contacts:
•
1 general failure alarm warning of furnace under temperature (20°C
below the set temperature), thermocouple separation, problem with RAM
backup after a re-set or adjustment error.
•
2 threshold alarms, with programmable activation direction and hysteresis.
The activation of a threshold alarm can be identi
fi
ed when the corresponding green LED
turns off on the Control Unit.
The activation of the general fault alarm can be identi
fi
ed when all the three green LEDs
turn off on the Control Unit.
The contacts are normally closed, dry and potential-free. The maximum switched load
is 10 W for each contact (up to 100 V or up to 0.5 A).
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...