XZR500 User’s Manual
Michell Instruments
21
OPERATION
3.3.2 Set-up [ - ]
The set-up mode is used for changing system control parameters, each of which are
selected from the set-up con
fi
guration table. There are two different sets of setting
parameters. Quick settings are accessible under code 0.12, while advance settings are
accessible under access code 0.20.
Access
Code
Function
Default
Setting
Unit
Remarks
0.12
Std con
fi
g. 0.30
0.00
N/A
Resets the system to standard con
fi
guration.
The default con
fi
guration is restored. The
current parameters will be deleted. Usually
necessary after replacement of the micro
controller card. Do not change without
consulting factory
Test gas value
8.00
%
Sets the value of the calibration gas
Type Alarm 1
2.00
N/A
Sets alarm type. If value >1.00 high Alarm 1,
if value <1.00 low Alarm 1
Level of Alarm 1
30.00
% of the
measurement range Sets process trigger point for Alarm 1
Type Alarm 2
2.00
Sets alarm type. If value >1.00 high Alarm 2,
if value <1.00 low Alarm 2
Level of Alarm 2
30.00
% of the
measurement range Sets process trigger point for Alarm 2
Hysteresis
1.00
% of the
measurement range Sets relay hysteresis
Fail safe value
0.00
% of the output
scale
Sets the fault alarm. If released, the output
signal switches to the selected value
0.20
Oven temp set
pt
700.00
°C
Sets the oven temperature. Do not change
without consulting factory
Francais
English
Italiano
English
N/A
Sets display language
Offset room
temp
3.00
N/A
Sets the offset set point for the MSRS T/C cold
junction. Do not change without consulting
factory
Room temp.
grad.
10.00
N/A
Sets the span set point. Do not change without
consulting factory
High adjust
0.00
Adjusts reading near 21%
0 or 4 - 20 mA
output
4.00
N/A
Sets the analog output signal
Rec. range low
0.01
%
Sets the measuring range lower limit
Rec. range high
10.00
%
Sets the measuring range higher limit
CO2 factor
0.00
%
Enter the CO2 value
Gr
10.00
Factory setting. Do not change without
consulting factory
Ti
10.00
Factory setting. Do not change without
consulting factory
Offset pressure
Pre-set
mbar
Sets the offset pressure. Do not change
without consulting factory
Absolute
pressure
Pre-set
mbar
Sets the absolute pressure. Do not change
without consulting factory
Cell cleaning
Start the cell self-cleaning process
Table 5
Access codes 0.12 & 0.20
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...