XZR500 User’s Manual
62
97137 Issue 8, April 2017
APPENDIX H
Appendix H
Second 4-20 mA Output
The optional 2nd analog output allows acquisition of a linear 4-20 mA current
corresponding to the 0.1% - 25% scale (this scale cannot be changed).
This 4-20 mA output is galvanic isolated (500 V).
Wiring of the 2nd 4-20 mA output on the connector (on the electronic card) is as follows:
•
terminal 27 = “-” 4-20 mA
•
terminal 28 = “+” 4-20 mA
h
-
+
27 28
Connector
Connector
29
38 39
50
Electronic Card
2nd 4-20 mA Output Connector
For the connection use
shielded cable 2 x 0.75mm².
Bare 8mm of the wire end on
the terminal’s side and
connect the braid to the
copper bar near the cable
gland.
Figure 26
Wiring of the Second 4-20 mA Output