XZR500 User’s Manual
12
97137 Issue 8, April 2017
INSTALLATION
4.
Place the gasket onto the inner tube (f) and insert into the probe head
(e) (see
Figure 8). Make sure the bevelled edge of the inner tube tip
faces the opposite direction to the gas inlet hole on the outer tube (c) as
shown below. There are three marks on the other end of the inner tube
to indicate the orientation of the bevelled edge (see
Figure 12).
Process flow
f. Inner tube
c. Outer tube
Bevelled edge
Gas inlet hole
Figure 12
Inner Tube Orientation
5.
Place the locking ring (h) with the locking ring gasket (g), the rear nut
(i) and the inner tube gasket (j) accordingly on the rear of the probe
head. (See
Figure 8)
6.
Put insulation between process and the probe head if necessary.
If the back
fl
ush option is chosen, then the orientation
must be reversed. This is because the inner tube becomes
the inlet so any dust build up will happen in the inner
tube. When the back
fl
ush is operated, it clears the tube
more ef
fi
ciently.