1.2 General
Remarks
The analyzer is made up of a measurement probe and a Control Unit.
1.2.1 Sensor Head and Probe
The semi in-situ arrangement consists of the following elements:
•
Sensor head, containing MSRS, sensor furnace, cable connection &
calibration port.
•
Probe, comprising of an inner and outer tube to allow
fl
ow of sample from
fl
ue to sensor.
The sample gas is returned to the
fl
ue practically unchanged in composition and
condition. This is due to the very small amount of sample required to diffuse into the
sensor furnace.
The gas
fl
ow is shown by the arrows in
Figure 4 below. In this example, the stack is
vertical and the
fl
ue gas
fl
ow direction is upwards.
Using the Pitot tube effect, gases enter through the hole near the tip of the outer tube
and circulate in the space between the outer tube (25, 2) and the inner tube (23).
During this process the gases contact the sensor through diffusion. They then
fl
ow into
the inner tube and to the
fl
ue via the bevelled edge.
The tubes are
fi
tted so that the bevelled edge of the inner tube and the holes of the
outer tube face in opposite directions.
8
25
6
2
1
3
4
5
7
9
10
11
12
13
14
15
16
17
18
19
20
21
24
22
23
26
27
28
29
1
Sensor housing
11
XZR500 connector block
20
M6 Nut for
fi
xing XZR500 probe plate
2
Outer tube
12
SS bulkhead union for cal.gas
21
M6 washer
3
Sealing head
13
2 sealing ferrules
22
M6 lock washer
4
XZR500 furnace
14
Nut for
fi
xing bulkhead union
23
Inner tube
5
XZR500 MSRS
15
Stainless steel plug
24
VITON 8 x 2.5 O-ring
6
Gasket for
fl
ange (x2)
16
M8 lock washer
25
Outer tube
7
Gasket for wheel
17
Cable gland
26
Locking ring
8
Inter-tubes gasket
18
M8 spacer
27
Gasket for plug
9
Gasket for sealing screw
19
CHC 6x25 screw for
fi
XZR500 sensor attachmt plate
28
Big rear nut
10
TRF 3x16 screw for connector
29
Inter-tube gasket
Figure 4
XZR500 Sensor Head & Probe
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...