INTRODUCTION
1.3 Speci
fi
cations
1.3.1 General
• Microcontroller:
Motorola 68HC12
• Inlet
converter:
analog, 16 bits resolution, 0.0015% linearity error,
with
fi
lter and embedded calibrator
• Outlet
converter:
analog, 12 bits resolution, ±1 bit linearity error
• Display
resolution:
0.1% O
2
(or 0.01 % on request and during
calibration)
•
The 3-button keypad and digital interface allow:
Reading
Continuous display of the oxygen concentration to one decimal place (option
of second dp). Other parameters available through the HMI are Furnace
Temperature, Thermocouple Junction Temperature, MSRS Voltage and
Ambient Pressure (in mbars), if the optional pressure sensor is ordered.
Con
fi
guration
Con
fi
guring of the test gas value, activation direction of the alarms (high
or low) and hysteresis level, the fail safe value (output value sent by the
electronics in case of general failure), language (English, French or Italian),
the output analog signal, the associated scale and the factory settings. (For
other parameters see speci
fi
c Appendix.)
• Output
signal:
0-20 mA or 4-20 mA (user-selectable) galvanic isolation
(500 V), linear and programmable, output range can be selected between
0.01% and 25% O
2
• Output
impedance:
> 1 k
Ω
• Alarms:
contacts are normally closed, dry and potential-free, the cutting
power is max. 10 W (up to 100 V or up to 0.5 A):
1.
General failure alarm warning of furnace under temperature (20°C
below the instruction), thermocouple separation, problems with
RAM backup after a re-set or adjustment error
2.
Threshold alarms, with programmable activation direction and
hysteresis
• Consumption:
110 VA
•
Ingress and impact protection:
Electronic enclosure: IP52 and IK05
Probe: IP53 and IK05
•
Storage temperature:
between -10 and +70°C
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...