![Michell Instruments XZR500ST User Manual Download Page 32](http://html1.mh-extra.com/html/michell-instruments/xzr500st/xzr500st_user-manual_1784650032.webp)
XZR500 User’s Manual
24
97137 Issue 8, April 2017
OPERATION
Alarm hysteresis
Relay hysteresis
0.10
This page shows the value for the relay hysteresis
that is expressed in percentage oxygen. Press
the [+] and [–] keys to adjust the value. Press
the [*]
Enter
key to con
fi
rm the selection and
move to the next parameter.
Fail safe value
Fail safe value
0.00
This page shows, in percentage, the fail-safe
value of the measurement range. This is the
percentage value that the 4-20 mA output will
go to under fault conditions. E.g. 100.00 means
the output will be 20mA (100%) under fault
conditions.
Press the [+] and [–] keys to adjust the value.
Press the [*]
Enter
key to con
fi
rm the selection.
The Access code page will appear.
3.3.2.2 Access code 0.20 - advanced settings
Access code?
0.20
At the Access code page, use the [–] and [+]
keys to enter the 0.20 code to access the
advance settings and press the [*]
Enter
key to
con
fi
rm the selection.
Oven temperature set-point
Oven temp set pt
700.00
This
fi
rst page under access code 0.20 is used
to set the temperature of the sensor oven.
CAUTION: Do not change the value without
consulting the factory.
Press the [*]
Enter
key to move to the next
parameter.
Languages
English ?
No * Yes -
English, French or Italian can be selected as the
display language. There are separate pages for
each language.
Press the [*]
Enter
key to stay with the current
language in use
French ?
Non * Oui -
Press the [–] key to change to the chosen
language
Italiano ?
No * Si -
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...