OPERATION
3.3.3.3 Calibration Procedure
Follow Sections 3.3.3.1 and 3.3.3.2 to connect the calibration gas to the analyzer (8%
oxygen bottle, if 1G or 2G is ordered). Do not switch on the calibration gas
fl
ow at this
time.
Your selection?
Visu * Cal + Set -
1.
From the main page, press and hold the
[*] (Enter key) until the main selection
menu appears.
Test gas value
8.00
2.
Press the [+] key from the main selection
menu to enter the calibration mode. The
fi
rst page displays the ‘Test gas value’.
Make sure that the displayed value is the
same as the calibration gas.
NOTE: This value is in %. For
example, 8.00 equals 8%.
If the displayed value is not the same value
as the calibration gas, exit the calibration
mode from next page (‘Inject gas’) and
then go to ‘Test gas value’ under ‘Set-
up menu’ Access code 0.12 to adjust it.
(Please refer to details in Section 3.3.2.1)
Press the [*] (Enter key) to move to the
next page.
Inject gas
Yes* No -
3. This page shows the calibration start
command.
[*] – Starts the calibration process (go to
step 4)
[–] key – Cancels the calibration process
and exits the calibration mode. If the ‘Test
gas value’ needs to be modi
fi
ed, then go
to ‘Test gas value’ under ‘Set-up menu’
Access code 0.12 to adjust it.
NOTE: Once the calibration process
starts (after pressing the [*] key),
the 4-20 mA output will be frozen
from this step until the end of the
purge procedure and will show the
last measured value.
4. Open the test gas cylinder to introduce
the gas into the analyzer. Adjust the
fl
ow
to the recommended level of 7 l/hour (±2
l/hour) and ensure that the
fl
ow is stable.
Higher or lower
fl
ow rates could affect
the accuracy of the calibration as well as
further measurements.
Summary of Contents for XZR500ST
Page 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Page 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Page 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Page 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Page 77: ......
Page 78: ...http www michell com...