Functional Description
4-Channel SPO Digital Oscilloscopes
49
Functional Description
III - TRIGGER
System (cont’d)
« Runt »
Trigger
page 1
Option
Settings
Description
Type
Runt
A trigger event occurs if the pulse (negative or positive)
crosses the first level (L1) but not the second level (L2)
of the window, before recrossing the (L1) level in a
given time.
Source
CH1, CH2, CH3, CH4
Select the trigger source CH1 or CH2 or CH3 or CH4
Polarity
Select the pulse polarity : Positive or Negative
Limit
Range
< , >
< > , > <
Select the limits for « Runt »
< Less than > Greater than
< > Outside the window > < Inside the window
Duration
« Runt » duration range : 2ns - 4.20s
Next Page
1/2
Open page 2/2
« Runt »
Trigger
page 2
Option
Settings
Description
Type
Runt
Time
Use the
"Universal"
knob to adjust the Runt
duration in the range
:
2ns to 10s.
Lower - Upper
Use the Level knob to adjust the Lower and
Upper levels 'L1' and "L2".
Coupling
DC AC LF Rej
HF Rej
Trigger source coupling.
Noise Reject
On Off
Enable or disable the « Noise Reject »
.
Next Page
2/2
Open page 1/2