Functional Description
4-Channel SPO Digital Oscilloscopes
17
Functional Description
AUTO SETUP
Auto Setup
«
Auto Setup
» button on « RUN CONTROL
” block
DOX3000 digital oscilloscopes have an « Auto Setup » function that
configures automatically the device to produce a display adapted to the
signals at the channel inputs.
After launching an “Auto Setup”, select, if necessary, the type of signal
displayed among the four proposed : Several periods, One period, Rising or
Falling edge at the center of the screen.
Auto Setup
Function
Option
Description
Multi-cycle
Autoset
and displays several cycle signal.
Single-cycle
Autoset
and displays single cycle signal.
Rising edge
Autoset
and displays the Rising edge.
Falling edge
Autoset
and displays the Falling edge.
Undo config.
The oscilloscope recalls the previous
Setup.
The “Auto Setup” sets the trigger source and displays the channels according
to the following criteria :
●
If multiple channels have an input signal, channel with the lowest frequency
signal has the priority.
●
If no channel is « On » and no signal is found, no channels are displayed.
●
If one or more channels are « On » and no signal is found, only the « On »
channels remains active.