Functional Description
46
4-Channel SPO Digital Oscilloscopes
Functional Description
III - TRIGGER
System (cont’d)
« Window » trigger
When the « window » trigger type is enabled, the oscilloscope triggers when the
signal exits, the window defined by the two levels L1 and L2, from the top or
bottom level.
Window Trigger
page 1
Option
Settings
Description
Type
Window
The window is defined by two voltage levels L1
and L2.
Source
CH1, CH2, CH3, CH4
Select the trigger source CH1 or CH2 or CH3 or CH4
Window
type
Absolute
Relative
Absolute:
the two levels are set independently
Relative:
we adjust the delta on either side of a
central value
.
Lower Upper
Center Delta
The trigger window is define by the lower and
upper value or by the delta relative to a central
value.
Coupling
DC AC
LF Reject
HF Reject
Trigger source coupling
Noise
Reject
ON
Off
Enable (On) or Disable(Off) the noise reject