Functional Description
4-Channel SPO Digital Oscilloscopes
129
IX - « Arbitrary Waveform Generator » (AWG) option
(cont‟d)
Automatic Calibration of the « Arbitrary Waveform Generator»
If the ambient temperature varies by more than 5°C and the instrument was left « On » for more than
30 minutes you can start an automatic calibration of the generator output.
1° Press the «
Wave Gen
» button to open the WAVEFORM menu
2° Press the «
Setting
» button to open the SETTING submenu
3° Press the «
AWG Self Cal
» button to launch the automatic calibration.
4° Press the «
Run/Stop
» button to exit the AWG self Cal
We can also set the generator output impedance (Output Load) :
High-Z
or
50Ω
To set the generator default values (Sine, 1kHz, 4Vpp, 0Vdc) press the «
Default
» button.