Functional Description
36
4-Channel SPO Digital Oscilloscopes
Functional Description
II - HORIZONTAL
System (cont’d)
HORIZONTAL pad
The
“HORIZONTAL pad” knobs allow to adjust the Time Base coefficient
« S/div » and the traces horizontal Position.
The two buttons allow to open the HORIZONTAL
menu (“
Horiz
”) and to activate
the “
Zoom
” function.
Horizontal
"POSITION" knob
1. Adjusts the trace horizontal position (the trigger position relative to the
center of the screen). The time resolution of this command depends on the
selected Time Base coefficient.
2. To reset to zero the horizontal position, push the
″Position″ knob.
S/div knob
1. Adjusts the Time Base coefficient
.
When the acquisition is stopped ( with
“RUN/STOP” button or in “SINGLE” mode), turn the « S/div » knob to expand
or compress the waveform.
2.
« S/div »
knob
adjusts the
coefficient of the « Main » or « Zoom » Time
Bases
.
In
Zoom
mode, changing the Time Base coefficient (S/div) modifies
the width of the « Zoom » window.
Horizontal Zoom
Use the horizontal
“Zoom” function to define the portion of the waveform to be
observed in details. The slowest « zoom » time base coefficient may not be
lower than the “main” time base coefficient.
Use the horizontal «
Position »
knob to move the « zoom window » position
and the «
S/div
»
knob to magnify or compress the “zoom window”.
"M"
stands for « Main » time base
, "Z"
stands for “Zoom” time base
.
The vertical
« blue »
arrows
indicate the horizontal position of the trigger
event in the Main time base and in the “Zoom” window.
The horizontal « yellow » arrow
indicate the vertical position of the trigger
level.
Note: The display of the « Trigger Position
» symbol in the “zoomed window” is
only possible if the position of the “zoomed window” includes the trigger event.
If the trigger event is outside the “zoomed window” the horizontal blue arrow
indicates the direction in which to move the “zoomed window” to display the
trigger.