Getting started
14
4-Channel SPO Digital Oscilloscopes
Getting started
Verification of
instrument
operation
To check the operation of the oscilloscope, perform the following steps:
Steps
1. Turn on the oscilloscope.
Press the
« Default »
button
of the « RUN CONTROL » pad to load the
« default » configuration of the oscilloscope. Note : The default value for the
attenuation compensation coefficient of the probe is 1X
.
2. Switch the attenuation of the
“probe” to the « 1X » position, plug it on CH1
and turn clockwise to block. Connect the probe tip and the ground alligator clip
to the terminals
“Cal 3V 1kHz” and “Ground” of the probe adjust output.
3. Press the "
Auto Setup
" button to display the
“3Vpp, 1kHz” Cal square wave.
4. Press the
"CH1"
button to disable CH1. Then press the "
CH2"
button to
enable
CH2 and repeat steps 2 and 3 to display the Cal signal on CH2
.
Repeat the previous steps with the channels CH3 and CH4
.
Cal Output : 3V 1kHz
GND
CH1 channel
«
Default Setup
»