Functional Description
4-Channel SPO Digital Oscilloscopes
19
Functional Description
I - VERTICAL System
Vertical knobs and buttons allow to display waveforms and to modify sensitivity and
vertical position.
Channels: CH1
CH2 CH3 CH4
page 1
Option
Value
Description
Coupling
DC
AC
GND
DC: passes the AC and DC components of the
input signal.
AC: blocks the DC component of the input signal
and attenuates AC components below 10Hz.
GND: disconnects the input signal.
Bandwidth
limit
ON
OFF
Limits the channel bandwidth to 20 MHz and
reduces high frequency noise.
Full bandwidth
Volts/Div
Coarse
Fine
Selects the resolution of the Volts/div knob :
« Coarse » :
defines a 1-2-5 sequence.
«
Fine
» : small steps continuous adjustement.
Probe
.1x, .2x, .5x,1x ,
5x, 10x, 20x,
50x,100x, 200x,
500x, 1000x,
2000x,
5000x,10000x
Set to match the attenuation of the probe you are
using to ensure correct vertical readouts.
Input
Impedance
1M
Ω
50Ω
Selects the channel input impedance.
«
1M
Ω
» : for use with
10MΩ input impedance
«1/10» probes.
«
50Ω
»
: for use with 50Ω cable and 50Ω output
impedance Generator to preserve impedance
matching which is necessary when working with
high frequency signals.
Next Page
Page 1/2
To access page 2/2
page 2
Option
Values
Description
Unit
V
A
Vertical unit :
V
olts or
A
mpères.
Invert
ON
OFF
To invert the channel.
To disable inversion.
Next Page
Page 2/2
To access page 1/2.
CH1 CH2 CH3 CH4
buttons to open vertical
menu and set channels
“on”
CH1 CH2 CH3 CH4
« Volt/div »
knobs
CH1 CH2 CH3 CH4
Vertical «
Position
» knobs