Functional Description
158
4-Channel SPO Digital Oscilloscopes
X - DECODE LIN Serial Bus Decode
(cont‟d)
Setting the LIN Trigger
The LIN trigger allows to trigger on the rising edge of the «
Sync Break
» signal (at the start of the
message frame), the frame identifier « ID » or the « ID+Data ».
1° Press the front panel «
Setup
» button to open the TRIGGER menu
2° Press the «
Type
» button and use the « Universal » to select Serial 1 or Serial 2 according to
Decode
3° Press the «
Trigger Setting
» button to open the
“LIN TRIG SET” submenu
Break
The oscilloscope triggers on a «
Sync Break exit
» (start of frame)
ID
The oscilloscope triggers on a specified frame ID
ID+Data
The oscilloscope triggers on a specified IDentifier and « Data »
Data Error
The oscilloscope triggers on a data error