MI 3325 MultiServicerXD
205
Parameters
State
On
– enables automatic settings of tickers in Visual and Functional tests.
Off
– disables automatic settings of tickers in Visual and Functional tests.
Operation after end of test
This flow command controls the proceeding of the Auto Sequence
®
in regard to the
measurement results.
Parameters
Operation after end of test
– pass
– fail
– no status
The operation can be individually set for the case the
measurement passed, failed or ended without a status.
Manual:
The test sequence stops and waits for
appropriate command (ENTER key, external
command...) to proceed.
Auto:
The test sequence automatically proceeds.
Result screen
This flow command controls the proceeding after the Auto Sequence
®
has ended.
Parameters
Auto Save
Auto Sequence
®
results are stored in the momentary
workspace.
A new Node with the date and time will be created. Under the
Node Auto Sequence
®
results or (if Appliance info flow
command is set) a new appliance and Auto Sequence
®
results will be stored.
Up to 100 Auto Sequence
®
results or appliances can be
automatically stored under the same node. If more results /
appliances are available, they are split to multiple nodes.
Local Save Flow setting is disabled by default.
Auto Print
Auto Sequence® results are automatically printed. If multiple
print options are possible Print label menu is opened before
printing,
Note
This flow command is active only if Auto Sequence
®
is started from the Auto Sequences
®
Main menu (not from Memory organizer).
E.6
Custom Inspections programming
Arbitrary set of tasks dedicated to specific user defined Inspections can be programmed with
application of Custom Inspection Editor Tool, accessible from Auto Sequence
® Editor
workspace. Custom Inspections are stored in dedicated file *.indf with user defined name. For