Available settings are organized in groups, each group starts with concerned Single test name.
Limit Calculator is referred to Continuity or PE_conductor(PRCD) functions. Refer to Single test
description chapter for details of parameters and limits setting / calculation.
Figure 7.5: Auto Sequence
®
Configurator menu
– opened from Auto Sequence
®
Main
menu on the left, on the right opened from Auto Sequence
®
view menu
Options
Starts the Auto Sequence® from Configurator menu.
Option is available, if configurator was opened from Auto Sequence
®
Main menu. New configuration is automatically applied to all related
single tests.
Confirms limits and parameters setting and returns to view menu.
Option is available, if configurator was opened from Auto Sequence
®
view
menu. Start the Auto Sequence
®
from view menu with confirmed
configuration.
7.2.1.4 Indication of Loops
The attached ‘x3’ at the end of single test name indicates that a loop of single tests is
programmed. This means that the marked single test will be carried out as many times as the
number behind the ‘x’ indicates. It is possible to exit the loop before, at the end of each
individual measurement.
7.2.1.5 Managing multiple points
If the device under test has more than one test point for an individual single test and the
selected Auto Sequence
®
predicts only one test point (one single test), it is possible to change
the Auto Sequence
®
appropriately. Single tests with enabled Multiple points ticker will be
executed in a continuous loop. It is possible to exit the loop anytime at the end of each
individual measurement.
The Multiple points setting is valid only for the actual Auto Sequence
®
. If the user often tests
appliances with more than one test point, it is recommended to programme a special Auto
Sequence
®
with pre-programmed loops.