MI 3325 MultiServicerXD
144
Connection diagram
Figure 6.126: Discharging time measurement
Measurement procedure
Enter the
Discharging time
function.
Set test parameters / limits.
Connect Residual voltage cable to the instrument and to the device under test (DUT),
see
Connect DUT to the mains supply and Switch it ON.
Start the measurement.
Measurement will stop automatically when disconnecting DUT from mains supply.
Save results (optional).
Figure 6.127: Discharging time test results
6.2.37 Functional inspection
Figure 6.128: Functional inspection start menu (left) and menu during inspection (right)