MI 3325 MultiServicerXD
165
11 Technical specifications
11.1
HV AC, HV AC programmable
Voltage a.c.
Range
Resolution
Accuracy
U
0 V ... 1999 V
1 V
(3 % of reading)
2.00 kV ... 5.99 kV
10 V
(3 % of reading)
Current a.c. (apparent)
Range
Resolution
Accuracy
I
0.0 mA ... 49.9 mA** / 99.9 mA*
0.1 mA
(3 % of r 3 D)
Current a.c.(capacitive, resistive)
Range
Resolution
Accuracy
I
r
0.0 mA ... 49.9 mA** / 99.9 mA*
0.1 mA
Indicative
I
c
-49.9 mA ... 49.9 mA**
-99.9 mA ... 99.9 mA*
0.1 mA
Indicative
* Output voltage 100 V…..2500 V
** Output voltage 2510
V……5000 V
Output voltage .................................................
100 V … 1000 V (-0/+10%),
........................................................................
1010 V … 5000 V (-0/+5%) floating to earth
Trip out time (if apparent current exceeds high limit).......... < 30 ms
Short circuit current ………………………………………….> 200 mA
O
utput power………………………………………………….250 VA max.
Test terminals:
All results
HV terminals
11.2
Continuity
Continuity resistance
Range
Resolution
Accuracy
R
0.00
… 19.99
0.01
(2 % of r 2 D)
20.0
… 99.9
0.1
3 % of reading
100.0
… 199.9
0.1
5 % of reading
200
… 999
1
indicative
Voltage drop ( Iout = 10 A)
Range
Resolution
Accuracy
ΔU
0.00 V ... 19.99 V
0.01 V
(2 % of r 5 D)
20.0 V ... 99.9 V
0.1 V
3 % of reading
Limit value of voltage drop versus wire cross-section:
Wire cross-section (mm
2
)
Limit voltage drop (V)
0.5
5.0