Test circuit
Figure 6.109: HV AC measurement
HV AC measurement procedure
Prepare test setup as mentioned in IMPORTANT SAFETY NOTES above.
Select the
HV AC
function.
Set test parameters / limits.
Connect HV test leads to HV terminals on the instrument.
Connect HV test leads to device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 6.110: Examples of HV AC measurement results
Note:
First HV measurement after power on the instrument (if password protection is enabled)
or first HV measurement after enabling or changing password require entering password
for enabling HV test. Refer to chapter
4.7.4 Change password for HV functions
more information.
6.2.32 HV AC programmable
IMPORTANT SAFETY NOTES
Refer to chapter
for more information regarding safe use of the
instrument.
Requirements from EN 50191 for test installations and safety of withstanding
voltage testing shall be applied. Prohibition zone is 30 mm and no part of the body