MI 3325 MultiServicerXD
Appendix B
185
B.4
Profile France (AUAC)
Modifications in chapters:
– Fault loop impedance and prospective fault current in system with RCD
6.2.21 Z auto - Auto test sequence for fast line and loop
650 mA added in the
I
N
parameter in
Test Parameters / Limits
section as follows:
I
N
Rated RCD residual current sensitivity
[10 mA, 15 mA, 30 mA, 100 mA, 300
mA, 500 mA, 650 mA, 1000 mA]
Nominal residual current (A, AC) ............ 10 mA, 15 mA, 30 mA, 100 mA, 300 mA, 500 mA,
650 mA, 1000 mA
RCD test current in relation to RCD type, nominal RCD current and multiplication factor
I
N
× 1/2
(mA)
I
N
× 1
(mA)
I
N
× 2
(mA)
I
N
× 5
(mA)
RCD I
I
N
(mA)
AC A, F
B,
B+
AC
A, F
B,
B+
AC
A, F
B,
B+
AC
A,
F
B,
B+
AC A,
F
B,
B+
10
5
3.5
5
10
20
20
20
40
40
50
100 100
15
7.5
5.3
7.5
15
30
30
30
60
60
75
150 150
30
15
10.5
15
30
42
60
60
84
120
150 212 300
100
50
35
50
100
141
200
200
282
400
500 707 1000
300
150 105
150
300
424
600
600
848
×
1500
×
×
500
250 175
250
500
707
1000 1000 1410
×
2500
×
×
650
325 227.5 250
650 916.5 1300 1300
×
×
×
×
×
1000
500 350
500 1000 1410
×
2000
×
×
×
×
×
×
×
. ............................................................ not applicable
. ............................................................ applicable
AC type................................................... sine wave test current
A, F types…… ........................................ pulsed current
B, B+ types ............................................. smooth DC current
Other technical specifications remain unchanged.
B.5
Profile UK (AUAB)
Setting options:
Option
Description
Isc factor
Short circuit current Isc in the supply system is important for selection or
verification of protective circuit breakers (fuses, over-current breaking
devices, RCDs). The value should be set according to local regulative.