MI 3325 MultiServicerXD
169
Total Harmonic Distortion (current)
Range
Resolution
Accuracy
THDI
0 mA
… 999 mA
1 mA
(5 % of r 5 D)
1.00 A ... 16.00 A
0.01 A
5 % of reading
Cos Φ
Range
Resolution
Accuracy
Cos Phi
0.00i
… 1.00i
0.00c
… 1.00c
0.01
(5 % of r 5 D)
Voltage
Range
Resolution
Accuracy
U
0.0 V
… 199.9 V
0.1 V
(3 % of r 10 D)
200 V
… 264 V
1 V
3 % of reading
Current
Range
Resolution
Accuracy
I
0 mA
… 999 mA
1 mA
(3 % of r 5 D)
1.00 A ... 16.00 A
0.01 A
3 % of reading
Accuracy is valid within 0.5c ≤ PF ≤ 0.8i
Test terminals:
P,S,Q,PF,THDU,THDI, Cos Phi, U,I
Socket L
– Socket N
11.9
Leak’s & Power
Power (active)
Range
Resolution
Accuracy
P
0.00 W…19.99 W
0.01 W
(5 % of r 5 D)
20.0 W…199.9 W
0.1 W
5 % of reading
200 W ... 1999 W
1 W
5 % of reading
2.00 kW ... 3.70 kW
10 W
5 % of reading
Power (apparent)
Range
Resolution
Accuracy
S
0.00 VA…19.99 VA
0.01 VA
(5 % of r 10 D)
20.0 VA…199.9 VA
0.1 VA
5 % of reading
200 VA ... 1999 VA
1 VA
5 % of reading
2.00 kVA...3.70 kVA
10 VA
5 % of reading
Power (reactive)
Range
Resolution
Accuracy
Q
0.00 VAr
…19.99 VAr
0.01 VAr
(5 % of r 10 D)
20.0 VAr
…199.9 var
0.1 VAr
5 % of reading
200 VAr ... 1999 var
1 VAr
5 % of reading
2.00 kVAr...3.70 kVAr
10 VAr
5 % of reading