MI 3325 MultiServicerXD
170
Power factor
Range
Resolution
Accuracy
PF
0.00i
… 1.00i
0.00c
… 1.00c
0.01
(5 % of r 5 D)
Total Harmonic Distortion (voltage)
Range
Resolution
Accuracy
THDU
0.0 %
… 99.9 %
0.1 %
(5 % of r 5 D)
Total Harmonic Distortion (current)
Range
Resolution
Accuracy
THDI
0 mA
… 999 mA
1 mA
(5 % of r 5 D)
1.00 A ... 16.00 A
0.01 A
(5 % of reading)
Cos Φ
Range
Resolution
Accuracy
Cos Phi
0.00i
… 1.00i
0.00c
… 1.00c
0.01
(5 % of r 5 D)
Voltage
Range
Resolution
Accuracy
U
0.0 V
… 199.9 V
0.1 V
(3 % of r 10 D)
200 V
… 264 V
1 V
3 % of reading
Current
Range
Resolution
Accuracy
I
0 mA
… 999 mA
1 mA
(3 % of r 5 D)
1.00 A ... 16.00 A
0.01 A
3 % of reading
Differential leakage current
Range
Resolution
Accuracy
Idiff
0.010 mA
… 1.999 mA
0.001 mA
(3 % of r 3 D)
2.00 mA
… 19.99 mA
0.01 mA
5 % of reading
Operating range (acc. to EN 61557-16) ........... 0.010 mA
… 19.99 mA
Influence of load current .................................. < 0.02 mA/A
Touch leakage current
Range
Resolution
Accuracy
Itou
0.010 mA
… 1.999 mA
0.001 mA
(3 % of r 3 D)
2.00 mA
… 19.99 mA
0.01 mA
5 % of reading
Operating range (acc. to EN 61557-16) ........... 0.010 mA
… 19.99 mA
Test terminals:
P,S,Q,PF,THDU,THDI, Cos Phi, U,I
Socket L
– Socket N
Idiff
Socket L,N
– Socket PE, P/S
Itou
Socket L,N
– P/S