MI 2792 PowerQ4 Plus
Operating the instrument
73
Connection
1W 4W 3W
Method of connecting the instrument to multi phase
systems (see 4.2.1 for details).
1W
: 1-phase 2-wire system;
3W
: 3-phase 3-wire system;
4W
: 3-phase 4-wire system.
Synchronization
Synchronization channel. This channel is used for
instrument synchronization to the network frequency.
Also a frequency measurement is performed on that
channel. Depending on
Connection
user can select:
1W
: U1 or I1.
3W
: U12, or I1.
4W
: U1, I1.
System frequency
Select system frequency.
50
Hz
60
Hz
Default parameters
Set factory default. These are:
Nominal range: 110 ÷ 240V (L-N);
Voltage ratio: 1:1;
Phase current clamps: Smart Clamps;
Neutral current clamps: Smart Clamps;
Connection: 4W;
Synchronization: U1
System frequency: 50 Hz.
Table 3.79: Keys functions
Change selected parameter value.
Select Connection setup parameter.
Enter into submenu.
Confirm Default parameters.
Back to the “MEASUREMENT SETUP” menu screen.