MI 2792 PowerQ4 Plus
Technical specifications
133
6.4.2 Compliance to the to the IEC 61000-4-30
IEC 61000-4-30 Section and
Parameter
PowerQ4
Plus
Parameter
Class
Measurement
Method -
IEC 61000-4-
30 Section
Uncertainty
Measuring
range
(1)
Influence
Quantity range
(2)
Aggregation
Method
(3)
5.1 Frequency
freq
S
5.1.1
± 10 mHz
10 Hz ÷ 70 Hz
40 Hz ÷ 70 Hz
Arithmetic
5.2 Magnitude of the Supply
U
Rms
S
5.2.1
± 0.5 % U
Nom
10 % ÷ 150 % U
Nom
10 % ÷ 150 % U
Nom
RMS
5.3 Flicker
P
st
S
5.3.1
± 5 %
(4)
0.4 ÷ 4.0
0 ÷ 10
IEC 61000-4-
15
5.4 Dips and Swells
U
Dip,
U
Swell
duration
S 5.4.1
± 0.5 %
± 1 cycle
> 10 % U
Nom
1.5 cycle ÷ 7 days
–
–
5.5 Interruptions
U
Int
duration
S 5.4.1
± 0.5 %
± 1 cycle
< 150 % U
Nom
1.5 cycle ÷ 7 days
– –
5.7 Unbalance
u
-
,
u
0
A
5.7.1
± 0.15 %
0.5 % ÷ 5 %
0 % ÷ 5 %
RMS
5.8 Voltage Harmonics
Uh
N
S
5.8.1
IEC 61000-4-7
Class II
0 % ÷ 20 % U
Nom
0 % ÷ 20 % U
Nom
RMS
5.9 Voltage Interharmonics
Uih
N
S
5.9.1
IEC 61000-4-7
Class II
0 % ÷ 20 % U
Nom
0 % ÷ 20 % U
Nom
RMS
5.10 Mains signalling voltage
U
Sig
S
5.10.2
± 5 % U
Sig
3 % ÷ 15 % U
Nom
0 % ÷ 20 % U
Nom
RMS
A.6.3 Magnitude of the current I
Rms
S
A.6.3.1
± 0.5 %
2 % ÷ 200 % I
Nom
2 % ÷ 200 % I
Nom
RMS
A.6.4 Inrush current
I
½Rms
S
A.6.4.1
± 1 %
2 % ÷ 200 % I
Nom
–
–
A.6.5 Harmonic currents
Ih
n
A
A.6.5
IEC 61000-4-7
Class II
0 % ÷ 200 % I
Nom
0 % ÷ 200 % I
Nom
RMS
A.6.6 Interharmonic currents
Iih
n
A
A.6.6
IEC 61000-4-7
Class II
0 % ÷ 200 % I
Nom
0 % ÷ 200 % I
Nom
RMS
(1) The instrument meets the uncertainty requirements for signals within the measuring range.
(2) The instrument tolerates signals in the influence quantity range without shifting the measurement of other parameters out of their uncertainty
requirement, and without instrument damage.
(3) RMS aggregation according to the IEC 61000-4-30 section 4.4 and 4.5, Arithmetic according to the section 5.1.14 in this manual.
(4) Guaranteed only in 49 ÷ 51Hz frequency range