MI 2792 PowerQ4 Plus
Operating the instrument
64
3 – event on phase U
3
12 – event on voltage U
12
23 – event on voltage U
23
31 – event on voltage U
31
Start
Event start time (when first U
Rms(1/2)
) value cross threshold.
T
Indicates type of event or transition:
D – Dip
I – Interrupt
S – Swell
Level
Minimal or maximal value in event U
Dip
, U
Int
, U
Swell
Duration Event
duration.
Table 3.67: Keys function
Group view is shown. Press to switch on “PHASE” view.
Phase view is shown. Press to switch on “GROUP” view.
Filter events by type:
Show all events
Show dips only
Show interrupts only
Show swells only
Filter events by phase:
Show only events on phase 1
Show only events on phase 2
Show only events on phase 3
Show all events
Show event summary (by types and phases):
Back to Group view.
Show details about the selected event: