MI 2883 Energy Master
Operating the instrument
66
(activate) General recorder.
General recorder is active, recording in progress
Note:
Recorder will run until one of the following end
conditions is met:
STOP
key was pressed by user
Given Duration criteria was met
Maximal record length was reached
SD CARD is full
Note:
If recorder start time is not explicitly given, recorder
start depends on Real Time clock multiple of interval. For
example: recorder is activated at 12:12 with 5 minute
interval. Recorder will actually start at 12:15.
Note:
If during record session instrument batteries are
drained, due to long interruption for example, instrument
will shut down automatically. After power restauration, it
will automatically start new recording session.
Interval
Select General recorder aggregation interval. The smaller
the interval is, more measurements will be used for the
same record duration.
Include events
Select whether events are included in the record.
On:
Record events signatures in table form (see
3.15 for details)
Off:
Events are not recorded
Include alarms
Select whether alarms are included in the record.
On:
Record alarm signatures in table form (see
3.16 for details)
Off:
Alarms are not recorded
Include signalling
events
Select whether signalling events according to the IEC
61000-4-30 should be included in the record.
On:
Signalling events included in the record
Off:
Signalling events are not recorded
Start time
Define start time of recording:
Manual, pressing function key F1
At the given time and date.
Duration
Define recording duration. General recorder will record
measurement for given time duration:
Manual,
1, 6 or 12 hours, or
1, 2, 3, 7, 15, 30, 60 days.
Recommended/maximal
record duration:
Show recommended and maximal Duration parameter for
giver recording Interval.
Available memory
Show SD card free space