MI 2883 Energy Master
Technical specifications
159
6.2.12
Energy
Measuring range
(kWh, kvarh, kVAh)
Resolution
Accuracy
Act
ive
e
n
e
rg
y
Ep
*
Excluding clamps
(Instrument only)
000,000,000.001 ÷ 999,999,999.999
12 digits
±0.5 %
Ep
With
A 1227, A 1446
Flex clamps
000,000,000.001 ÷ 999,999,999.999
±2.0 %
Ep
With A 1281
Multirange
clamps 1000 A
000,000,000.001 ÷ 999,999,999.999
±1.0 %
Ep
With A 1033
1000 A
000,000,000.001 ÷ 999,999,999.999
±2.0 %
Ep
R
e
a
ct
ive
e
n
e
rg
y
Eq
**
Excluding clamps
(Instrument only)
000,000,000.001 ÷ 999,999,999.999
12 digits
±0.5 %
Eq
With
A 1227, A 1446
Flex clamps
000,000,000.001 ÷ 999,999,999.999
±2.0 %
Eq
With A 1281
Multirange
clamps 1000 A
000,000,000.001 ÷ 999,999,999.999
±1.0 %
Eq
With A 1033
1000 A
000,000,000.001 ÷ 999,999,999.999
±2.0 %
Eq
*Accuracy values are valid if cos
φ
0.80, I
10 % I
Nom
and U
80 % U
Nom
**Accuracy values are valid if sin
φ
0.50, I
10 % I
Nom
and U
80 % U
Nom
6.2.13
Voltage harmonics and THD
Measuring range
Resolution
Accuracy
Uh
N
< 3 % U
Nom
10 mV
± 0.15 %
U
Nom
3 % U
Nom
< Uh
N
< 20 % U
Nom
10 mV
± 5 %
Uh
N
U
Nom
: Nominal voltage (RMS)
Uh
N
: measured harmonic voltage
N
:
harmonic component 0
th
÷ 50
th
Measuring range
Resolution
Accuracy
0 % U
Nom
< THD
U
< 20 % U
Nom
0.1 %
± 0.4
U
Nom
: nominal voltage (RMS)
6.2.14
Current harmonics, THD and k-factor
Measuring range
Resolution
Accuracy
Ih
N
< 10 % I
N
om
10 mV
± 0.15 %
I
Nom
10 % I
N
om
< Ih
N
< 100 %
10 mV
± 5 %
Ih
N
I
N
om
: Nominal clamp current (RMS)
Ih
N
:
measured harmonic current
N
:
harmonic component 0
th
÷ 50
th