MI 2883 Energy Master
Operating the instrument
53
Figure 3.35: Flickers trend screen
Table 3.32: Instrument screen symbols and abbreviations
Pst1m1,
Pst1m2,
Pst1m3,
Pst1m12,
Pst1m23,
Pst1m31
Maximal ( ), average ( ) and minimal ( ) value of 1-minute short
term flicker P
st(1min)
for phase voltages U
1
, U
2
, U
3
or line voltages
U
12
, U
23
, U
31
Pst1,
Pst2,
Pst3,
Pst12,
Pst23,
Pst31
Maximal ( ), average ( ) and minimal ( ) value of 10-minutes
short term flicker P
st
for phase voltages U
1
, U
2
, U
3
or line voltages
U
12
, U
23
, U
31
Plt1,
Plt2,
Plt3,
Plt12,
Plt23,
Plt31
Maximal ( ), average ( ) and minimal ( ) value of 2-hours long
term flicker P
lt
in phase voltages U
1
, U
2
, U
3
or line voltages U
12
,
U
23
, U
31