Chapter 9 – Errors
52
Bad drying results
Apart from a proper device function, the drying process depends decisively
on the correct setting up and loading of the autoclave.
What you can do
Check the correct set-up of the autoclave. If necessary increase the
tilt by further screwing out the front feet of the unit.
The floor of the vessel must be free. Where applicable, remove fallen
instruments, filter paper, etc.
The tank filter is clogged. If necessary, check and clean it.
Pay attention to the correct loading of the autoclave (see page 20
Load the autoclave)
Do not overload the autoclave. Pay attention that textiles have no
direct contact with the chamber wall and floor.
Activate the pre-heating (see page 23,
Select automatic pre-heating
)
Employ the function Additional Drying (see page 24,
Additional
Drying
)