Chapter 9 – Errors
49
Error message
Possible causes
What you can do
Error 1:
Vacuum system
Door seal, sealing surface at the
chamber is polluted or defective
Check door seal and sealing surface at the
chamber for contamination and possibly clean
Check door seal for defects, possibly switch,
see page 11
Door seal wrongly inserted
Check door seal for correct seating
Error 2:
Steam generator
autoclave is overloaded
Decreased heating power, since
supply voltage too low
Pay attention to the correct loading quantities
(see page 20Load the autoclave ,
Load the
autoclave
)
Check building-side electrical Installation,
operate device on a trial basis on another
electric circuit.
If problem persists, notify specialist dealers
Error 4:
Pressure release
Waste water hose is bent or
installed with a large droop
Check correct laying of the drain hose from
the cooling water (with continuous gradient
and kink-free)
Error 6:
Ventilation
The sterile filter is polluted,
beforehand appropriate warning
text appears
Exchange sterile filter
Error 8:
Time-base
Maximum difference between
the program expiring date and
the internal computer clock is
exceeded
On repeated occurrence, notify specialist
dealers
Error 9:
Door open
The sliding closure grip was
pushed upwards during a
running program
Push the sliding closure grip downwards until
the stop. Correct display:
Door closed
On repeated occurrence notify your specialist
dealer
Error 10:
Overheat Steam
generator
Capillary tube regulator "level
control" is opened at program
start (error message immediately
after start) or during a program
run (until the end of the
sterilization) the monitoring time
the capillary tube regulator is
switched back (with water feed
from feed water) is exceeded
After program termination and immediate
start, this error message can be repeated
after a pause time of two minutes
On repeated occurrence notify your specialist
dealer
Error 12:
Door locking
Locking pin of the door is stiff
Check the smooth running of the locking pin
On repeated occurrence notify your specialist
dealer
Error 13:
No cooling water
See warning text
Warning no cooling
water
Error 14:
No feed water
see warning text
Warning no feed water
,
however this error appears after the start of a
program
Error 18:
Sensor: ...Input: … The device-internal inspection of
the sensors for temperature,
pressure or conductivity resulted
in a too large deviation, the
message can occur when the
device is switched on or during a
program
On repeated occurrence notify your specialist
dealer