Chapter 5 – Logging
34
program. The display shows
Printout
.
After finishing the printout the display again shows
Print stored
cycles
.
in case you would like to leave the submenu without printing, press
the K
EY
(S).
NOTICE:
A termination
during
the printing is only possible by disconnecting the
instrument at the mains switch or interrupting the power supply of the
printer.
In order to leave the menu, likewise press the K
EY
(S). The display
shows the setup menu
Function: Print
.
By again pressing the K
EY
(S), you completely leave the menu and
return to the starting position of the display.
Delete stored logs
In order e.g. to suppress the warning text
Printer memory full
for
preset option
Immediate printout No
, the saved logs can be
manually deleted.
In the following example you can read how you can delete all saved logs.
Program
Start-Stop
Select the SETUP Menu
Function
by short simultaneous pressing
of the keys (+) and (-).
The display shows
Function: Print.
Press the K
EY
(P). The Menu
Print – File transmission
is
displayed.
Navigate with the K
EY
(+) or (-), until the display shows
Delete all cycles
.
Press the K
EY
(P) in order to delete all logs now.
In order to cancel the submenu without deleting, press the K
EY
(S).
In order to leave the menu after deletion, press the K
EY
(P). The
display shows
Function: Print
.
By again pressing the K
EY
(S), you completely leave the menu and
return to the starting position of the display.
Display printer memory
Program
Start-Stop
With a connected and initialised printer you can verify how many logs are
already saved in the log memory of the autoclave.
Proceed as follows:
Select the SETUP Menu
Function
by short simultaneous pressing
of the keys (+) and (-).
The display shows
Function: Print
.
Navigate with the K
EY
(+) or (-), up to the display for the Storage
Occupancy (see left Display Figure) are shown.
Press the K
EY
(P), in order in to reach the submenu
Print - File
Transmission
.
With the K
EY
(+) or (-) navigate until you see the current storage
occupancy on the display.
In order to leave the menu again, press the K
EY
(S) twice.