Chapter 3 – Initial start-up
In this chapter you learn
Who may set up, install and start up the autoclave
Which prerequisites you must create for setting up, installing and starting up the
autoclave
Conditions for setting up, installing and
starting up
■
Only have the autoclave set up, installed, and started up by people
→
authorized by MELAG.
Danger!
■
According to the currently valid VDE standards, this device is not
suitable for use in areas subject to explosion hazards.
■
The autoclave is intended for use outside the patient environment.
The minimum distance to the treatment area must have a radius of at
least 1.5 metres.
Damages to the autoclave and/or injuries can occur in case this
precaution is not observed.
Set-up location requirements
Location
Set up the autoclave in a dry and dustproof location. The air humidity
should be 30% to 60% and the ambient temperature 16° to 26 °C.
Warning!
■
Without fail, keep the specified distance free to the surrounding
surfaces at the sides and above the autoclave.
Non-observance can lead to heat accumulation. This could
adversely affect the function of the autoclave and entail the
shortening of the vacuum pump's service life and result in extended
program times.
Distance to the surrounding
surfaces
The distance above and to the side of the autoclave to the surrounding
surfaces must be at least five centimetres.