Chapter 9 – Errors
46
Before you call
Follow the operating procedures shown in connection with a warning text
or an error message on the
→
display of the autoclave. You will find the
most important events in the following table. Possible causes and
appropriate operating instructions are listed for the events.
If you do not find the event in the table below or your efforts do not lead to
success, turn to your specialist dealer or the
→
authorised MELAG
customer service centre in your vicinity. In order to help you, keep the
serial number of your autoclave and a detailed fault description of the fault
report ready.
Warning message
Possible causes
What you can do
Warning: Door open /
No start possible
Door switch is not closed on start Push the sliding closure grip until the stop
downwards. Correct display:
Door closed
Warning: No cooling
water/ Check tap water
Faucet not opened
water pressure provided by the
customer too low
protective motor switch has
triggered
vacuum pump seized after longer
idle time
Check cooling water inlet, open faucet
Check cooling water pressure, (install increase in
pressure, e.g. house waterworks)
Press the reset button of the motor circuit breaker
(Fig. 1 Reset button of overheating protection)
again
Turn vacuum pump (see Technical Manual)
Warning: No feed
water/
check aqua dem/dest
quality
Too little or no water in the
storage tank
Intake hose bent
Suction level too high (max.
suction height 1.5 m from the
bottom of the external water
storage tank to the intake stub of
the autoclave
Suction filter of the external water
storage tank clogged
Feed water supply from external storage tank
Check filling level, possibly top up of the feed
water
Check kink-free laying of the intake hose
Check the position of the suction nozzle on the
floor of the water storage tank
Clean suction filter
Feed water-supply from
MELAdem
®
47
Check water treatment unit, possibly open inflow
to the unit. When the pressure reservoir is empty
about 1 h after restart and the message is
repeated, have the water treatment unit checked
by the MELAG customer service
Notice!
At the first start-up or re-startup, this
message can occur once since the piping system
is still not completely filled with feed water: restart