Chapter 9 – Errors
48
Warning message
Possible causes
What you can do
Electric power supply of the printer
is interrupted
Ensure electric power supply, the red LED "P" at
the log printer MELA
®
42 must light red
Printer is "offline"
Set printer to "online" (press key "SEL" at the
MELA
prin
t
®
42, LED "SEL" must light green)
Printer memory full
The device-internal log memory is
allocated (max. 40 logs possible).
The message is shown at the start of a program.
Repeatedly press the K
EY
(S) to delete the
message, and the program starts. The oldest low
is then deleted.
A printer/ computer/ CF Card
Writer is registered and in the
menu
Function: Print
the
option
Immediate printout
No
is set.
Set printer to
Immediate printout Yes
(see
page 32,
Automatic immediate output logfiles
Delete printer memory (see page 34,
Delete
stored logfiles
), possibly print all saved logs
beforehand (see page 34,
Print all stored
logfiles
Release the printer in the menu
File
transmission
and set the option
No printer
Execute Service
The servicing message is
activated, the device has reached
the prescribed
→
batches or
number of years has expired
The message is shown at program start.
Repeatedly press the K
EY
(S) to delete the
message, and the program starts.
Carry out maintenance with MELAG customer
service or service of the specialist trade
NOTICE:
Service counter is reset by the customer
service
Test not successful
Leakage rate: 3.2
The leakage rate determined by
the vacuum test lies above the
maximum permissible value of
1.3 mbar.
Door seal, chamber flange is
contaminated
Check door seal and chamber flange for
cleanliness, possibly clean
Check door seal for damage, possibly exchange,
see page 37,
Chapter 6 – Maintenance
Door seal wrongly inserted
Check door seal for correct seating
Repeat vacuum test with completely cold device
Warning! Battery dead
The monitoring of the device's
internal battery voltage has
determined a too low voltage
value.
The battery must be exchanged by MELAG
customer service or service of the specialist trade