Chapter 5 – Logging
35
Read log files correctly
Header
The head of the program listing includes the general fundamental
information for the completed program such as for example current date,
the selected program, the daily batch number and the autoclave type.
Values of the program steps
During the program the program sequence phases are recorded with the
accompanying values for steam pressure, temperature and time (relative
to the program start).
Summary
The summary states whether the program was successfully concluded.
Furthermore the values of the required sterilization time, the sterilization
temperature and the pressure including its maximum deviations are
displayed.