Chapter 9 – Errors
51
Error message
Possible causes
What you can do
temperature
Check door seal for wear
On repeated occurrence notify your specialist
dealer
Error 36:
Sterilization PU
Shortfall of the minimum
sterilization pressure
Carry out vacuum test
Operate device with smaller load
On repeated occurrence notify your specialist
dealer
Error 37:
Sterilization PO
Exceeding the maximum
permissible sterilization pressure
On repeated occurrence notify your specialist
dealer
Error 38:
Sterilization TD
Maximum permitted difference
between measured and
theoretical temperature was
exceeded
On repeated occurrence notify your specialist
dealer
No display on the screen
After switching on the autoclave the display remains empty.
What you can do
Check whether the power plug is correctly in the socket.
Check the supply voltage at the socket.
If necessary, switch the instrument securing at the lower front of the
autoclave (see page10,
Fig. 1/(8)
)off. Follow the instructions in the
technical manual in
Exchange Instrument Securing
.
Too large feed water consumption
The feed water consumption is dependent upon the program and the
loading of the autoclave.
What you can do
Check the correct set-up of the autoclave. If necessary, adjust the tilt of
the autoclave by further screwing out the front feet of the unit.
The floor of the vessel must be free. Where applicable, remove fallen
instruments, filter paper, etc.