Chapter 3 – Initial start-up
16
Electrical connection
Provide the following electrical connections for the autoclave:
Electric circuit with 230 V and 50 Hz
16 A separate fuse protection – at least Automat Type B
Protection from leakage current 30 mA
Connected load
Vacuklav
®
24 B+
2500 W
Vacuklav
®
30 B+ : 2400 W
Cooling water inlet
A pressure hose is provided for connecting the cooling water inlet to a
customer-side faucet.
NOTICE:
In order to prevent water damage, MELAG recommends the use of a leak
monitor, e.g. water stop valve from MELAG (Art. No. 01056)
Cooling water discharge
The waste-water hose can be connected as follows:
On an existing rinsing drain when using the dual-chamber siphon
(Art. No. 26635) instead of the existing siphon
On a separate drain pipe when using the surface-mounted siphon
from MELAG (Art. No. 37410 – not in the standard scope of delivery)
The drain must be located below the autoclave.
NOTICE:
The waste-water hose must be laid dip-free with continuous descent.
For waste-water paths which are longer than two metres we recommend a
fixed laying of HT (high temperature) piping by a specialised installation
firm.
Observe the following precautionary measures when handling the
mains cable and power plug:
Danger of injury!
■
Never splice or modify the mains cable.
■
Never bend or twist the mains cable.
■
Never pull on the mains cable to take it out of the socket. Always
take hold of the power plug itself.
■
Never place any heavy objects on the mains cable.
■
Never run the mains cable over places where it could be squeezed
(e.g. doors or windows).
■
Do not run the mains cable along a heat source.
■
Do not use any nails, staples or similar objects to fixate a cable.
If the mains cable or the power plug is damaged, turn off the autoclave.
Mains cables or power plugs may only be replaced by
→
authorised
persons.
If these points are not observed, the cable or plug can be damaged
and/or a fire or electrical shock might occur. Serious injuries could
be the consequence.