Humidity
90% RH at 40 °C (104 °F) max.
Calibration temperature
20 °C (68 °F)
Maximum altitude
3000 m (9843 ft.)
IP rating
IP 54
Physical
Display
Full LCD colour screen with user configurable back-light
Dimensions
228 x 105 x 75 mm (8.98 x 4.1 x 2.95 in)
Weight
0,93 kg (2,05 lbs)
Fuse
x2 500 mA (FF) 1000 V 32 x 6 mm ceramic fuse, high break capacity HBC,
30 kA minimum Glass fuses must not be installed.
Operational Uncertainty,
IEC
61557
Insulation
Fiducial Value A
E
1
E
2
E
3
Operational Uncertainty
0.1 MΩ to 0.99 MΩ 0.1 MΩ
22 %
0
0
0
22 %
1 MΩ to 9.99 MΩ
1 MΩ
4 %
0
0
0
4 %
10 MΩ to 99.9 MΩ
10 MΩ
4 %
0
0
0
4 %
100 MΩ to 999 MΩ 950 MΩ
2.2 % 0
0
5.4 % 8.5 %
Continuity
Fiducial Value A
E
1
E
2
E
3
Operational Uncertainty
0.1 Ω to 2 Ω
0.1 Ω
23 %
0
0
0
23 %
Key to table:
A: Intrinsic uncertainty
E
1
: Reference position ±90º
E
2
: Supply Voltage
E
3
: Temperature 0 ºC to 30 ºC
IEC61557 -1, 61557-2 and 61557-4
Specifications
82
www.megger.com
Summary of Contents for MTR105
Page 1: ...Baker Instruments MTR105 Rotating Machine Tester User Guide ...
Page 7: ... www megger com vii ...
Page 92: ... 86 www megger com ...
Page 93: ... www megger com 87 ...
Page 94: ... 88 www megger com ...
Page 95: ... www megger com 89 ...
Page 96: ... 90 www megger com ...
Page 97: ... www megger com 91 ...
Page 98: ... 92 www megger com ...
Page 99: ... www megger com 93 ...