Caution
•
For the communication protocol implementation guide, contact MagnaService.
•
This implementation should be performed on a network with the characteristics described in 2.4.1 by an IT
specialist.
•
IT network failures to provide the required characteristics may lead to delays in data communication or
incorrect, incomplete or corrupted data transmission, resulting in incorrect information to the user.
•
Only rely on the ventilator for alarm signal generation and information. The IT network is not
reliable for receiving alarm signals. Therefore, do not use a distributed alarm system as the
only means to recognize alarm signal generation.
The parameters are monitored based on the pressure and the FiO2 measurements performed by oxygen monitor.
Figure 5: Left side view
Table 5: Description of the components found in the left side
•
Only use power supply, parts, pieces and accessories specified by MAGNAMED listed in this
manual, which were tested and approved for use in conjunction with this equipment;
otherwise, this can jeopardize the operation endangering the patient or user.
Description of the items in Figure 5
19. OXYGEN INLET
Connect oxygen. Inlet pressure should be in the range of 39 to 87 psi (270 to
600 kPa). Standard DISS connection (ABNT NBR-11906:1992).
20. SERIAL CONNECTION
Serial Communication
–
RS-232
–
female DB-9 connector for software
update.
21. INLET +12VDC
–
External AC/DC Source
Power supply input +12VDC
–
Connection to external AC/DC source through
the plug
22. On/Off Switch
On/Off Switch
23. Air Inlet Filter
Ambient air intake filter which is mixed with oxygen to provide oxygen
concentrations less than 100%. Filter should be replaced according to the
descriptive operations manual (12.Replacing the Ambient Air Filter).