40
OxyMag_rev26
Caution
•
Never sterilize or immerse the CO2 sensor in liquid;
•
Do not apply voltage to the sensor cable;
•
Do not use the CO2 sensor in environments which specifications are outside the limits stablished in the
Technical Specifications (Temperature, Humidity, etc.)
•
Airway adapters of the CO2 sensor are non-sterile accessories. Steam sterilization can damage these
accessories.
•
Never sterilize or immerse the CO2 sensor in liquid.
6.7
Technical Specifications of the Capnography
Table 15: Technical Specifications of the Capnography
Attribute
Specification
GENERAL
Description
Mainstream monitoring sensor with infrared technology.
Dimensions (W x D x H)
IRMA CO
2
: 38 x 37 x 34mm (1.49” x 1.45” x 1.34”)
Cable length
2.50m (± 0.02m)
Weight
< 25g (without cable);
< 38g (with cable).
Mechanical Resistance
Withstands repeated 1-m falls on a hard surface.
According to the standard requirements for ambulances (EN
1789:2004
–
clause 6.4) and requirements against shock and vibration
(ISO 80601-2-55 - ed.1).
Electric Power Supply
IRMA CO
2
: 4.5
–
5.5 VDC, Max 1.0W (power measured with 5V and
less than 350mA during 200ms).
Surface Temperature (room temperature 23°C)
IRMA CO
2
:
Max: 41°C / 106°F.
Airway adapter
Adult/Pediatric (Disposable):