99
OxyMag_rev26
Specifications of the Monitoring Ventilation Parameters
Table 31:
Ventilation parameters
Item
Parameter
Range
Resolution
Tolerance
Unit
1
Instant Pressure
Measured
-20 to 100
1
(1 cmH2O or
2% reading)
cmH
2
O
(2)
2
Maximum Inspiratory
Pressure
0 to 90
1
(1 cmH2O or
2% reading)
cmH
2
O
3
Measured Pressure
0 to 90
1
(1 cmH2O or
2% reading)
cmH
2
O
4
Plateau Pressure
0 to 90
1
(1 cmH2O or
2% reading)
cmH
2
O
5
PEEP
- Pressure at the end
of expiration
-20 to 90
1
(1 cmH2O or
2% reading)
cmH
2
O
6
Intrinsic PEEP at the
end of expiration
-20 to 90
1
(1 cmH2O or
2% reading)
cmH
2
O
7
Flow Measured
(Sensor Adult)
(9)
-150 to 150
1
± (2.0L.min-1 or
5% reading)
L.min
-1
8
Flow Measured
(Sensor Pediatric)
(9)
-50 to 50
0,5
± (0.5L.min-1 or
5% reading)
L.min
-1
9
Flow Measured
(Sensor Neonatal)
(9)
-20 to 20
0,2
± (0.2L.min-1 or
5% reading)
L.min
-1
10
Volume Measured
(Sensor Adult
–
ADU)
(3) (9)
100 to 3000
100 to 995:5
(20ml or 5%
value measured)
mL
1000 to
3000:10
11
Volume Measured
(Sensor Pediatric
–
INF)
(3) (9)
10 to 400
2
(10ml or 5%
value measured)
mL
12
Volume Measured
(Sensor Neonatal
–
NEO)
(3) (9)
1 to 100
1
(3ml or 5%
value measured)
mL
13
Minute volume
(Sensor Adult
–
ADU)
(9)
0,1 to 99,0
0,001
(0.18L or 3%
value
measured)
(4)
L