23
OxyMag_rev26
4.2
Noninvasive ventilation mask
Use of the breathing circuit for NON INVASIVE VENTILATION (VNI or NIV
–
Non Invasive Ventilation).
A.
Without HME filter;
Figure 6: Assembly of the non-invasive mask without HME filter
B.
With mask and HME filter
Figure 7: Assembly of non-invasive mask with HME filter
•
Use HME Filter and MASKs specified by
MAGNAMED use appropriate MASKs for
each type of patient.
•
Position correctly the diaphragm and
the exhalation valve to prevent
obstruction of the expiratory limb;
•
The correct connection of the pressure
outlet tubes and the absence of
obstruction are extremely important for
the proper functioning of the patient’s
ventilation monitoring and, therefore, it
should be checked frequently during the
course of ventilation of the patients.
•
Never obstructe pressure port.The
measured pressure in this points are
used by the patient ventilation
monitoring system.
•
All connections must be FIRMLY
secured to prevent leakage.
•
Only use parts, pieces and accessories
specified by MAGNAMED listed in this
manual, which have been tested and
approved for use in conjunction with
this equipment; otherwise, it can
jeopardize the operation endangering
the patient or user;
•
PROPER connection of these pressure
line tubes is extremely important for
monitoring patient ventilation.
•
When using Oxymag for extended time
in battery, an alarm occurs which