OxyMag_rev26
Table 8: List of values adopted by the equipment when selecting a patient
After startup, it is possible to change the height value within the adjustment range of the patient type set by clicking on the chart
area and menu and selecting the Settings button (General Tab). Setting is as table below:
Table 9: List of adjustment range for height and weight
Patient Type
Height Adjustment [m]
Ideal Weight
P [Kg]
Min.
Máx.
NEONATAL
0,16
0,52
≤ 6,0
PEDIATRIC
0,53
1,08
6,0 < P ≤ 25
ADULT
1,09
2,5
> 25
The weight of the patient considered by the equipment is the ideal body weight, calculated according the height of the patient.
Height adjustment of the patient does not remain after turning off the equipment. It is only possible to change the height within the range
of values corresponding to the type of patient selected.
•
USE THE FLOW SENSOR INDICATED. The correct ventilation monitoring depends on the flow
sensor used in the breathing circuit.
•
Even if there is need to use breathing circuits different from the patients to be ventilated, the
FLOW SENSOR SHOULD BE AS INDICATED.
Startup button
Patient type
Height [m]
Ideal Weight P [kg]
NEONATAL
0,36
2,8
PEDIATRIC
0,95
19,8
ADULT
1,50
49,5